VEGA VEGAPULS 66 (≥ 2.0.0 - ≤ 3.8) standpipe ver. 4 … 20 mA_HART two-wire User Manual

Page 48

Advertising
background image

Process conditions

F

or the process temperature and the vessel pressure, you also have to note the specifications on

the type label! Always the lowest value is applicable!

P

rocess temperature (measured on the process fitting depending on process seal

-

FKM (V

iton)

-40 … +80 °C (-40 … +176 °F)

-

FKM (V

iton) with temperature adapter

-40 … +150 °C (-40 … +302 °F)

-

FFKM (K

alrez 6375)

-20 … +150 °C (-4 … +302 °F)

-

graphite (coupling cone ceramic)

-60 … +250 °C (-76 … +482 °F)

-

graphite (coupling cone ceramic) with
temperature adapter

-60 … +400 °C (-76 … +752 °F)

V

essel pressure relating to the antenna cone

-

PTFE

-100 … 4000

kPa/-1 … 40 bar (-14.5 … 580 psi)

with PN 40

-

C

eramic

-100 … 16000

kPa/-1 … 160 bar (-14.5 … 2321 psi)

with PN 160

V

essel pressure relating to the flange nom-

inal stage

see supplementary instructions manual "Flanges
according to DIN-EN-ASME-JIS
"

V

ibration resistance

mechanical vibrations with 1 g and 5 … 100 Hz

8)

8)

T

ested according to the regulations of German Lloyd, GL directive 2.