31

9 Supplement

VEGASWING 63 • - NAMUR

29231-EN-130409

Density

0.7 … 2.5 g/cm³ (0.025 … 0.09 lbs/in³); 0.5 … 2.5 g/cm³

(0.018 … 0.09 lbs/in³) by switching over

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

1)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Screw terminals

for wire cross-section up to 1.5 mm² (AWG 16)

Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry

Ʋ Single chamber housing

– 1 x IP 68 cable gland M20 x 1.5; 1 x blind plug

M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Connection cable

Ʋ Wire cross-section

0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

Ʋ Tensile strength

< 1200 N (270 lbf)

Ʋ Standard length

5 m (16.4 ft)

Ʋ Max. length

1000 m (3280 ft)

Ʋ Min. bending radius

25 mm (0.984 in) with 25 °C (77 °F)

Ʋ Diameter approx.

8 mm (0.315 in)

Ʋ Colour - standard PE

Black

Ʋ Colour - standard PUR

Blue

Ʋ Colour - Ex-version

Blue

Adjustment elements
Mode switch

Ʋ Min.

rising characteristic curve (High current when immersed)

Ʋ Max.

falling characteristics (Low current when immersed)

Sensitivity switch

Ʋ 0.5

0.5 … 2.5 g/cm³ (0.018 … 0.9 oz/in³)

Ʋ 0.7

0.7 … 2.5 g/cm³ (0.025 … 0.9 oz/in³)

Test key

simulation of a line break between sensor and process-

ing unit

1)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.