VEGA VEGAPULS 68 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual
Page 13
1
3
2
100%
0%
4
Fig. 3: Measuring range (operating range), max. measuring distance and
reference plane
1
full
2
empty (max. measuring distance)
3
Measuring range
4
Reference plane
Information:
If the medium reaches the antenna, buildup can form on it and cause
faulty measurements later on.
The emitted radar impulses of VEGAPULS 68 are electromagnetic
waves. They have hence an electric and a magnetic share which are in
rectangular relation. The polariation plane is defined by the direction of
the electric share. With radar sensors, the polarisation can be used to
reduce the effect of false echoes considerably by turning the
instrument in the connection flange or mounting boss. The position of
the polarisation plane is marked on the instrument.
Polarisation plane
VEGAPULS
68 • Foundation Fieldbus
13
4 Mounting
29264
-EN