VEGA VEGAFLEX 63 Foundation Fieldbus User Manual

Page 46

Advertising
background image

Influence of the ambient temperature to the sensor electronics
Temperature drift

0.03 %/10 K relating to the max. measuring range
or max. 0.3 %

Temperature drift - Digital output

3 mm/10 K relating to the max. measuring range or
max. 10 mm

Ambient conditions

Ambient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

Process pressure

-

Flange version ≤ 2"/DN 50

-0.5 … +16 bar/-50 … +1600 kPa (-7.3 … +232 psig),
depending on the process fitting

-

Flange version > 2"/DN 50

-0.2 … +16 bar/-20 … +1600 kPa (-2.9 … +232 psig),
depending on the process fitting

Process temperature (flange temperature)

-40 … +150 °C (-40 … +302 °F)

The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.

0 °C

(32 °F)

40 °C

(104 °F)

20 °C

(58 °F)

60 °C

(140 °F)

80 °C

(176 °F)

50 °C

(122 °F)

-40 °C

(-40 °F)

100 °C

(212 °F)

150 °C

(302 °F)

1

2

Fig. 25: Ambient temperature - Process temperature

1

Ambient temperature

2

Process temperature (depending on the seal material)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

4)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x closing cap M20 x 1.5; 1 x blind stopper
M

20 x 1.5

4)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.

46

VEGAFLEX

63 • Foundation Fieldbus

10 Supplement

31844

-EN

-100426

Advertising