VEGA VEGAFLEX 67 (-40…+150°C PFA insulated) Foundation Fieldbus User Manual

Page 40

Advertising
background image

-

Rod: ø 10 mm (0.394 in)

approx. 620 g/m (6.7 oz/ft)

-

Cable: ø 4 mm (0.157 in)

approx. 80 g/m (0.86 oz/ft)

-

Gravity weight

325 g (11.5 oz)

Insulation thickness

-

Rod version

2 mm (0.08 in)

-

Cable version

0.75 mm (0.03 in)
On the gravity weight and beginning of the cable:
2 mm (0.08 in)

Probe length L (from seal surface), cannot be shortened

-

Rod: ø 10 mm (0.394 in)

up to 4 m (13.12 ft)

-

Trimming accuracy - rod

< 1

mm (0.039 in)

-

Cable: ø 4 mm (0.157 in)

up to 32 m (105 ft)

-

Trimming accuracy - cable

±0

.05 %

Lateral load - rod: ø 10 mm (0.394 in)

4 Nm (3 lbf ft)

Max. tensile load with cable: ø 4 mm
(0.157 in)

2 KN (450 lbf)

Input variable

Measured variable

Level of liquids

Min. dielectric figure (lower medium)

by 10 higher than the ε

r

of the upper medium

Min. layer thickness (upper medium)

> 100

mm (3.937 in)

The measured value deviates from the actual level
until the min. layer thickness is reached.

Max. layer thickness (upper medium)

-

Dielectric figure (upper medium) 10

6 m (19.69 ft)

-

Dielectric figure (upper medium) 5

10 m (32.81 in)

-

Dielectric figure (upper medium) 2

20 m (65.61 in)

Min. dielectric figure (upper medium) with
rod, cable version

ε

r

> 1

.6

40

VEGAFLEX

67 • Foundation Fieldbus

10 Supplement

32306

-EN

-120329

Advertising