VEGA VEGAFLEX 67 (-40…+150°C PFA insulated) Foundation Fieldbus User Manual
Page 50
Diagram, adjustment
The following illustration shows the function of the adjustment.
Secondary_value_1
Calibration
Lowest Point
Calibration
Highest Point
Sensor_value
cal_level_lo
cal_level_hi
Calibration
Lowest Point
Calibration
Highest Point
Sensor reference plane
cal_level_lo [%]
cal_level_hi [%]
Fig. 24: Adjustment VEGAFLEX 67
Parameter list
The following list contains the most important parameters and their meaning:
l
primary_value
-
This is the process value after adjustment and Linearization with the status of the transducer
block
-
It is the input for the AIFB when CHANNEL = 1 is selected
-
The unit is defined in "primary_value_unit"
l
primary_value_unit
-
Selected unit code for "primary_value"
l
secondary_value_1
-
This is the process value after adjustment with the status of the transducer block
-
It is the input for the AIFB when CHANNEL = 2 is selected
-
The unit is defined in "secondary_value_2_unit"
l
secondary_value_1_unit
-
Selected unit code for "secondary_value_1"
50
VEGAFLEX
67 • Foundation Fieldbus
10 Supplement
32306
-EN
-120329