VEGA VEGAFLEX 67 (-40…+150°C) 4 … 20 mA_HART two-wire User Manual

Page 27

Advertising
background image

1

M

ove from the measured value display to the main menu by

pushing [OK].

B

asic adjustment

D

isplay

D

iagnostics

S

ervice

I

nfo

2

S

elect the menu item "Basic adjustment" with [->] and confirm with

[OK]

. N

ow the menu item "Dielectric value" is displayed.

D

ielectric value

Upper medium

000

3

E

nter the dielectric figure (DK value) of the upper medium and

confirm with [OK].

S

ave the settings with [OK] and move to "Min. adjustment" with [-

>

]

.

M

in. adjustment

0

.00 %

=

10

.000 m(d)

8.000

m(d)

4

P

repare the % value for editing with [OK] and set the cursor to the

requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.

5

E

nter the suitable distance value in m for the empty vessel (e.g.

distance from the sensor to the vessel bottom) corresponding to
the percentage value.

6

S

ave the settings with [OK] and move to "Max. adjustment" with [-

>

]

.

P

roceed as follows:

M

ax. adjustment

100

.00 %

=

1

.000 m(d)

2.000

m(d)

1

P

repare the % value for editing with [OK] and set the cursor to the

requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.

2

E

nter the appropriate distance value in m (corresponding to the

percentage value) for the full vessel. Keep in mind that the max.
level must lie below the dead band.

Carry out max. adjust-
ment

VEGAFLEX

67 • 4 … 20

mA/HART - two-wire

27

6 S

et up with the indicating and adjustment module PLICSCOM

31853

-

EN

-120329

Advertising