4 software update – VEGA VEGAFLEX 67 (-40…+150°C) 4 … 20 mA_HART two-wire User Manual

Page 41

Advertising
background image

I

n Ex applications only one instrument and one electronics module with

respective Ex approval may be used.

I

f there is no electronics module available on site, one can be ordered

from the VEGA agency serving you.

T

he new electronics module must be loaded with the settings of the

sensor. These are the options:

l

A

t the factory by VEGA

l

O

r on site by the user

I

n both cases, the sensor serial number is necessary. The serial

numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
W

hen loading on site, first of all the order data must be downloaded

from the Internet (see operating instructions manual "Electronics
module
").

T

he electronics modules are adapted to the respective sensor and

distinguish also in the signal output or power supply.

E

lectronics module FX-E.67H suitable for VEGAFLEX 67 - 4 … 20 mA/

HART:

l

FX-E.67HX (X =

without approvals)

l

FX-E.67HC (C =

approvals XM, CX, CM, CK, CI, DX, DM, DI

according to product list)

8

.4 Software update

T

he software version of VEGAFLEX 67 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

Y

ou can view all software histories on our website www.vega.com.

M

ake use of this advantage and get registered for update information

via e-mail.

T

he following components are required to update the sensor software:

l

S

ensor

l

V

oltage supply

l

VEGACONNECT

l

PC

with PACTware

l

C

urrent sensor software as file

Sensor serial number

Assignment

4

… 20 mA/HART

VEGAFLEX

67 • 4 … 20

mA/HART - two-wire

41

8 M

aintenance and fault rectification

31853

-

EN

-120329

Advertising