ake note of the corresponding installation regulations for Ex

applications.

5

.2 Connection procedure

P

roceed as follows:

1

W

ire the connection cable up to the connection compart-

ment. The bending radius must be at least 25 mm.

2)

2

C

onnect the wire ends to the screw terminals according to

the wiring plan

P

roceed as follows:

1

S

nap VEGABOX 02 onto the carrier rail or screw it to the

mounting plate

2

L

oosen the cover screws and remove the cover

3

S

hift cable through the cable gland into VEGABOX 01 B

4

L

oosen the screws with a screwdriver

5

I

nsert the wire ends into the open terminals according to

the wiring plan

6

T

ighten the screws with a screwdriver

7

C

heck the hold of the wires in the terminals by lightly

pulling on them

8

T

ighten the compression nut of the cable entry. The seal

ring must completely encircle the cable

9

C

onnect the supply cable according to steps 3 to 8

10 S

crew the housing cover on

T

he electrical connection is finished.

2)

T

he connection cable is already preconfectioned. After shortening the ca-

ble, fasten the type plate with support again to the cable.

Select connec-
tion cable for Ex
applications

Direct connection

Connection via
VEGABOX 02

18

VEGAWELL

51 • 4 … 20

mA/HART

C

onnecting to power supply

32468

-EN