Fig. 8: Flange mounting of the radar sensor

4.3 Mounting instructions

Mount VEGAPULS 67 at least 200 mm (7.874 in) away from the vessel
wall.

If you cannot keep this distance you should carry out a false echo
storage before setup. This applies mainly if buildup on the vessel wall
is expected. In this case, we recommend repeating a false echo
storage later with existing buildup.

A corresponding collar flange for DN 80 (ASME 3" or JIS 80) as well as
a suitable adapter flange are available for mounting VEGAPULS 67.

With the housing versions plastic, aluminium single chamber and
stainless steel, the collar flange can be placed directly over the
housing. With the aluminium double chamber housing, retroactive
mounting in this way is not possible - the mounting type must be
specified with the order.

Information:
The socket should be as short as possible and the socket end should
be rounded. Interfering reflections from the vessel socket are thus
minimised.

Mounting position

Socket

VEGAPULS

67 • Profibus PA