VEGA VEGAPULS 67 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual

Page 52

Advertising
background image

Measuring range

up to 15 m (49.21 ft)

Reference conditions to measuring accuracy (similar to DIN EN 60770-1)

Reference conditions according to DIN EN 61298-1

-

Temperature

+

18 … +30 °C (+64 … +86 °F)

-

Relative humidity

45 … 75 %

-

Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Other reference conditions

-

Reflector

ideal reflector, e.g. metal plate 2 x 2 m

-

False reflections

Largest false echo 20 dB smaller than the useful
echo

Characteristics and performance data

Frequency

K-band

Interval

approx. 1 s

Beam angle -3 dB

7)

10°

Step response or adjustment time

8)

>

1 s (dependent on the parameter setting)

Max. level change

Adjustable up to 1 m/min. (dependent on the
parameter settings)

Max. emitted HF power of the antenna system

-

Pulse peak power

<

10 mW

-

Pulse duration

<

2 ns

-

Average power

<

25 µW

-

Average power with 1 m distance

<

1 µW/cm

2

Measuring accuracy

Resolution, general

max. 1 mm (0.039 in)

Deviation

9)

see diagrams

7)

Corresponds to the range with 50 % of the emitted power

8)

Time to output the correct level (with max. 10 % deviation) after a sudden
level change.

9)

Incl. non-linearity, hysteresis and non-repeatability.