VEGA VEGAWAVE 63 - NAMUR User Manual

Page 32

Advertising
background image

1

2

3

-50°C

(-58°F)

50°C

(122°F)

40°C

(104°F)

-40°C

(-40°F)

80°C

(176°F)

0°C

(32°F)

100°C

(212°F)

150°C

(302°F)

200°C

(392°F)

250°C

(482°F)

Fig. 32: Ambient temperature - Process temperature

1

Process temperature

2

Ambient temperature

3

Temperature range with temperature adapter

Product density

-

Standard

>

0.02 g/cm³ (0.0007 lbs/in³)

-

adjustable

>

0.008 g/cm³ (0.0003 lbs/in³)

Granular size

max. 10 mm (0.4 in)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

1)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5

Spring-loaded terminals

for wire cross-section up to 1.5 mm² (AWG 16)

Electromechanical data - version IP 66/IP 68 (1 bar)

Cable entry

-

Single chamber housing

l

1 x IP 68 cable gland M20 x 1.5; 1 x blind
stopper M20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Connection cable

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

<

0.036 Ω/m (0.011 Ω/ft)

1)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.

32