If one of the stated malfunctions is detected or in
case of power failure, the electronics takes on a
defined switching condition, i.e. the relay deener-
gises (safe condition).

Functional principle
The tuning fork is piezoelectrically energised and
vibrates at its mechanical resonance frequency of
approx. 150 Hz. When the tuning fork is sub-
merged in the product, the vibration amplitude
changes. This change is detected by the integrated
oscillator and converted into a switching command.

2

.3

Storage and transport

Packaging
Your instrument was protected by packaging
during transport. Its capacity to handle normal
loads during transport is assured by a test
according to DIN EN 24180.

The packaging consists of environment-friendly,
recyclable cardboard. In addition, the sensor is
provided with a protective cover of cardboard.
Dispose of the packaging material via specialised
recycling companies.

3 M

ounting

3

.1

General instructions

Switching point
In general, VEGAWAVE S 61 can be installed in
any position. The instrument simply has be
mounted in such a way that the vibrating element is
at the height of the desired switching point.

Moisture

Fig. 8: Measures against moisture penetration

Transport
Do not hold VEGAWAVE S 61 on the vibrating
element. Remove the protective cover just before
installation.

Handling
The vibrating level switch is a measuring instru-
ment and must be treated accordingly. Bending the
vibrating element will destroy the instrument.

Warning:
The housing must not be used to screw the
instrument in! Applying tightening force can
damage internal parts of the housing.

Use the hexagon above the thread for
screwing in.

3

.2

Installation instructions

Socket
The vibrating element should protrude into the
vessel to avoid buildup.

Filling opening
Mount the instrument in such a way that the tuning
fork does not protrude directly into the filling
stream.

Horizontal mounting

20°

a.

b.

Fig. 9: Horizontal mounting

a

Convex mounting

b

Concave mounting

14