5 Connecting to voltage supply

5.1 Preparing the connection

Generally note the following safety instructions:

l

Connect only in the complete absence of line voltage

l

If overvoltage surges are expected, overvoltage arresters

should be installed

In hazardous areas you should take note of the appropriate

regulations, conformity and type approval certificates of the

sensors and power supply units.

The voltage supply of the module card is provide via the

common power supply unit of the VEGALOG system.
You can find detailed information on the power supply in

chapter "Technical data" in the "Supplement".

5.2 Connection cable

Power supply is connected with standard cable according to

the national installation standards.
To connect 4 … 20 mA instruments, standard two-wire cable

without screen can be used. If electromagnetic interference is

expected which is above the test values of EN 61326 for

industrial areas, screened cable should be used.

Connect the cable screen on both ends to ground potential.

The screen must be connected (low impedance) on the

4 … 20 mA instrument to the potential equalisation.
If potential equalisation currents are expected, the screen

connection on the side of VEGALOG 571 EA must be made

via a ceramic capacitor (e.g. 1 nF, 1500 V). The low frequency

potential equalisation currents are thus suppressed, but the

protective effect against high frequency interference signals

remains.

Take note of the corresponding installation regulations for Ex

applications. In particular, make sure that no potential equal-

isation currents flow over the cable screen. In case of

grounding on both sides this can be achieved by the use of a

capacitor or a separate potential equalisation.
The total capacitance of the cable and of all capacitors must

not exceed 10 nF in Ex applications.

Note safety instructions

Take note of safety

instructions for Ex

applications

Select power supply

Selecting connection cable

Cable screening and ground-

ing

Select connection

cable for Ex applica-

tions

12

VEGALOG 571 EA - Input card analogue

Connecting to voltage supply