6 saving the parameter adjustment data – VEGA VEGABAR 63 Profibus PA User Manual

Page 43

Advertising
background image

6.6 Saving the parameter adjustment data

It is recommended noting the adjusted data, e.g. in this operating
instructions manual and archive them afterwards. They are hence
available for multiple use or service purposes.

If VEGABAR 63 is equipped with an indicating and adjustment module,
the most important data can be read out of the sensor into indicating
and adjustment module. The procedure is described in the operating
instructions manual "Indicating and adjustment module" in the menu
item "Copy sensor data". The data remain there permanently even if
the sensor power supply fails.

If it is necessary to exchange the sensor, the indicating and adjustment
module is inserted into the replacement instrument and the data are
written into the sensor under the menu item "Copy sensor data".

VEGABAR

63 • Profibus PA

43

6 Set up with the indicating and adjustment module PLICSCOM

32463

-

EN

-081209

Advertising