3 exchange or shorten cable/rod – VEGA VEGAFLEX 62 4 … 20 mA_HART two-wire User Manual

Page 45

Advertising
background image

8

.3 Exchange or shorten cable/rod

T

he cable or rod (meas. part) of the probe can be shortened, if

necessary. To loosen the meas. part you need two fork spanners with
spanner width 8.

1

L

oosen the measuring component by applying a fork spanner to

the flat surfaces (SW 8), provide counterforce with another fork
spanner (SW 8)

2

U

nscrew the loosened measuring component manually

3

P

lace the enclosed new double washer onto the thread.

Caution:
M

ake sure that the two components of the double washer remain

together.

4

S

crew in a new measuring part manually

5

E

xert counterforce with the second fork spanner and tighten the

measuring component on the flat surfaces with a torque of 15 Nm
(11

lbf ft).

Fig. 27: Exchanging the cable or rod

Information:
P

lease keep the stated torque so that the max. tensile strength

remains.

T

he cable and rod of the probe can be shortened by any amount.

1

C

able: loosen the three pins on the gravity weight (hexagon 4)

2

C

able: remove the pins

Exchanging the cable/
rod

Shorten cable/rod

VEGAFLEX

62 • 4 … 20

mA/HART two-wire

45

8 M

aintenance and fault rectification

31837

-

EN

-100426

Advertising