VEGA VEGAFLEX 62 4 … 20 mA_HART two-wire User Manual

Page 51

Advertising
background image

-

T

rimming accuracy - rod

<

1

mm (0.039 in)

-

C

able: ø 6 mm (0.236 in)

up to 60 m (196.9 ft)

-

T

rimming accuracy - cable

±

0.05 %

L

ateral load with rod: ø 16 mm (0.63 in)

30 N

m (22 lbf ft)

M

ax. tensile load with cable: ø 6 mm

(0.236

in)

30 KN (6745

lbf)

T

hread in gravity weight (cable version)

M

12

Input variable

M

easured variable

L

evel of liquids and solids

M

in. dielectric figure of the medium

ε

r

>

1.6

2

4

5

3

1

2

4

5

3

1

Fig. 29: Measuring ranges of VEGAFLEX 62

1

Reference plane

2

Probe length L

3

Measuring range (default setting refers to the measuring range in water)

4

Upper dead band (see diagrams under Accuracy - grey section)

5

Lower dead band (see diagrams under Accuracy - grey section)

Output variable

O

utput signal

4 … 20

mA/HART

C

ycle time

min. 1 s (dependent on the parameter setting)

S

ignal resolution

1.6 µA

VEGAFLEX

62 • 4 … 20

mA/HART two-wire

51

10 S

upplement

31837

-

EN

-100426