VEGA VEGAMIP R61 Receiving unit - Transistor User Manual

Page 29

Advertising
background image

Level

Switching status

Signal lamp -
Switching output
(yellow)

Mode min.
Dry run protection

open

Failure of the supply
voltage
(max./min. mode)

any

open

Malfunction

any

open

Red failure LED

lights

Depending on the switching function your output should carry out, you
can determine the switching direction by means of the slide switch.

l

For maximum indication or overfill protection set the slide switch to
position max.

l

For minimum indication or dry run protection, set the slide switch to
position min.

The transistor output is closed when the yellow signal lamp lights.

With the two keys you can change the LED indication strip and adjust
the switching point to the medium.

The keys are lowered to avoid unintentional changes of the setting. For
adjustment, use for example, a small screwdriver.

If you press one of the keys for more than two seconds, the adjustment
range shifts automatically in the respective direction. The greater the
distance between emitting and receiving unit, the greater the adjust-
ment range of the indication. The LED indication strip always shows a
small section of the actual measuring range.

Pressing the "<--" key makes the sensor more sensitive. Pressing the
"

-->" key makes the sensor less sensitive.

Sensitivity adjustment

VEGAMIP R

61 • - Transistor

29

6 Setup

40694

-

EN

-120228

Advertising