4 instructions for installation – VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA User Manual

Page 16

Advertising
background image

16

4 Mounting

VEGAPULS 61 • Profibus PA

36501-EN-121011

Fig. 6: Adjustment of the angle of inclination

Fig. 7: Turning by fastening in the centre

4.4 Instructions for installation

For tight installation of the version with plastic horn antenna with com-

pression or adapter flange, the following conditions must be fulfilled:
1. Use suitable flat seal, e.g. of EPDM with Shore hardness 25 or 50
2. Make sure the number of flange screws corresponds to the num-

ber of flange holes

3. Tighten all screws with the torque stated in the technical data

The emitted radar impulses of the radar sensor are electromagnetic

waves. The polarisation plane is the direction of the electrical wave

component. By turning the instrument in the connection flange or

mounting boss, the polarisation can be used to reduce the effects of

false echoes.
The position of the polarisation plane is marked on the process fitting

of the instrument.

1

2

Fig. 8: Position of the polarisation level
1 Marking with version with encapsulated antenna system

2 Marking with version with plastic horn antenna

When mounting the sensor, keep a distance of at least 200 mm

(7.874 in) to the vessel wall. If the sensor is installed in the center of

Tight installation of the

plastic horn antenna

Polarisation plane

Installation position

Advertising