VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA User Manual
Page 64
64
11 Supplement
VEGAPULS 61 • Profibus PA
36501-EN-121011
Max. torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing
10 Nm (7.376 lbf ft)
Ʋ Aluminium/Stainless steel housing
50 Nm (36.88 lbf ft)
Input variable
Measured variable
The measured quantity is the distance between process
fitting of the sensor and product surface. The reference
plane is the seal surface on the hexagon or the lower
side of the flange.
3
4 2
1
Fig. 55: Data of the input variable
1 Reference plane
2 Measured variable, max. measuring range
3 Antenna length
4 Useful measuring range
Standard electronics
Max. measuring range
35 m (114.8 ft)
Recommended measuring range
Ʋ Encapsulated antenna system
up to 10 m (32.81 ft)
Ʋ Plastic horn antenna
up to 20 m (65.62 ft)
Electronics with increased sensitivity
Max. measuring range
35 m (114.8 ft)
Recommended measuring range
Ʋ Encapsulated antenna system
up to 10 m (32.81 ft)
Ʋ Plastic horn antenna
up to 20 m (65.62 ft)
Output variable
Output signal
digital output signal, format according to IEEE-754
Sensor address
126 (default setting)
Damping (63 % of the input variable)
0 … 999 s, adjustable
Profibus PA profile
3.02