6

Setup

6

.1 Setup

Setup is carried out according to the operating instructions manual of
the respective sensor.

The measuring system can detect the following statuses (situations):

l

No liquid present

l

Water or hydrous (conductive) liquid present

l

Oil or a liquid similar to oil (non-conductive) present

l

Oil layer on water present

In general, liquids from a height of approximately 25 mm (1 in) can be
detected.

Oil layers up to a thickness of max. 50 mm (2 in) can be detected.
Upwards of this layer thickness, the measuring system detects only
the oil layer.

Note:
We recommend connecting the level switches in such a way that the
switching circuit is open when there is a level signal, line break or
failure (safe condition).

An open switch symbol corresponds to the detection of a liquid in the
following drawing

10

Float for oil/water detection • for VEGASWING 63 and EL 3

6 Setup

36676