VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) HART - integrated accumulator User Manual

Page 36

Advertising
background image

36

6 Set up with the indicating and adjustment module

VEGAPULS 61 • HART - integrated accumulator

40792-EN-121011

Caution:

If a separation of liquids with different dielectric figure occurs in the

vessel, for example through condensation, the radar sensor can

detect under certain circumstances only the medium with the higher

dielectric figure. Keep in mind that layer interfaces can cause faulty

measurements.
If you want to measure the total height of both liquids reliably, please

contact our service department or use an instrument specially de-

signed for interface measurement.
The following options are available when "Bulk solid" is selected:

The following features form the basis of the applications:
Silo (slim and high):

Vessel of metal: weld joints

Process/measurement conditions:

– Filling too close to the sensor

– System noise with completely empty silo increased

– Automatic false signal suppression with partly filled vessel

Bunker (large-volumed):

Vessel of concrete or metal:

– Structured vessel walls

– Installations present

Process/measurement conditions:

– Large distance to the medium

– Large angles of repose

Bunker with fast filling:

Vessel of concrete or metal, also multiple chamber silo:

– Structured vessel walls

– Installations present

Process/measurement conditions:

– Measured value jumps, e.g. by truck loading

– Large distance to the medium

– Large angles of repose

Heap:

Sensor mounting on movable conveyor belts

Detection of the heap profile

Height detection during filling

Process/measurement conditions:

– Measured value jumps, e.g. by the profile of the heap or trav-

erses

– Large angles of repose

– Measurement near the filling stream

Crusher:

Vessel: installations, wear and protective facilities available

Advertising