3 exchanging the electronics module – VEGA VEGAFLEX 66 (-200…+400°C) Profibus PA User Manual

Page 41

Advertising
background image

Status accor-
ding to NE 107

Error code

Text message

Cause/Rectification

no measured value available

Sensor does not find an echo, e.g.
due to faulty installation or wrong
parameter adjustment

no measured value available

Wrong sensor length entered

E

017

Adjustment span too small

Adjustment not within the specifica-
tion. Carry out the adjustment again,
increasing the distance between
min. and max. adjustment

E

036

No operable software

Failed or interrupted software up-
date/Repeat software update

E

042

Hardware error, electronics defec-
tive

Exchange the instrument or send it
in for repair

E

043

Hardware error, electronics defec-
tive

Exchange the instrument or send it
in for repair

E

113

Communication conflict

Exchange the instrument or send it
in for repair

Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.

In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.

If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.

The new electronics module must be loaded with the settings of the
sensor. These are the options:

l

At the factory by VEGA

l

Or on site by the user

In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Oscillator").

Reaction after fault rec-
tification

Sensor serial number

VEGAFLEX

66 • Profibus PA

41

8 Maintenance and fault rectification

34175

-

EN

-100426