VEGA VEGAFLEX 66 (-200…+400°C) Foundation Fieldbus User Manual

Page 46

Advertising
background image

2

4

3

5

1

2

4

5

3

1

Fig. 20: Measuring ranges of the VEGAFLEX 66 - rod and cable version

1

Reference plane

2

Probe length

3

Measuring range

4

Upper dead band (see diagrams under Accuracy - grey section)

5

Lower dead band (see diagrams under Accuracy - grey section)

Output variable

Output

-

Signal

digital output signal, Foundation Fieldbus protocol

-

Physical layer

according to IEC 61158-2

Cycle time

min. 1 s (dependent on the parameter setting)

-

Damping (63 % of the input variable)

0 … 999 s, adjustable

-

Fulfilled NAMUR recommendations

NE

43

Channel Numbers

46

VEGAFLEX

66 • Foundation Fieldbus

10 Supplement

34176

-EN

-100426

Advertising