VEGA VEGAFLEX 66 (-200…+400°C) Foundation Fieldbus User Manual
Page 46
Advertising
2
4
3
5
1
2
4
5
3
1
Fig. 20: Measuring ranges of the VEGAFLEX 66 - rod and cable version
1
Reference plane
2
Probe length
3
Measuring range
4
Upper dead band (see diagrams under Accuracy - grey section)
5
Lower dead band (see diagrams under Accuracy - grey section)
Output variable
Output
-
Signal
digital output signal, Foundation Fieldbus protocol
-
Physical layer
according to IEC 61158-2
Cycle time
min. 1 s (dependent on the parameter setting)
-
Damping (63 % of the input variable)
0 … 999 s, adjustable
-
Fulfilled NAMUR recommendations
NE
43
Channel Numbers
46
VEGAFLEX
66 • Foundation Fieldbus
10 Supplement
34176
-EN
-100426
Advertising