Fig. 27: Deviation VEGAFLEX 66 in coax version in water

1

Dead zone - no measurement possible in this area

L

Probe length

L

0

-3 mm (-0.118

")

3 mm (0.118

")

-10 mm (-0.394

")

10 mm (0.394

")

0,1 m (3.937

"

)

0,15 m

(1.968

")

0,1 m

(3.937

")

0,25 m (9.842

"

)

1

1

Fig. 28: Deviation VEGAFLEX 66 in coax version in oil

1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

Temperature drift

0.03 %/10 K relating to the max. measuring range
or max. 0.3 %

Temperature drift - Digital output

3 mm/10 K relating to the max. measuring range or
max. 10 mm

Ambient conditions

Ambient, storage and transport temperature

-

40 … +80 °C (-40 … +176 °F)

VEGAFLEX

66 • Foundation Fieldbus

51

10 Supplement