8 Maintenance and fault rectification

VEGABAR 67 • Foundation Fieldbus

36744-EN-130321

F

total

: Total deviation

F

perf

: Basic accuracy

F

stab

: Long-term drift

F

T

: Temperature coefficient (influence of medium or ambient

temperature)

F

Kl

: Deviation

F

a

: Error current output

Level measurement 1500 mmWs
Product temperature 40 °C, reference temperature 20 °C
VEGABAR 67 with measuring range 0.2 bar
Deviation < 0.1 %
Calculation ΔT: ΔT = 40 °C - 20 °C = 20 K
Calculation of the set Turn Down:
TD = 200 mbar/147 mbar, TD = 1.4
Basic accuracy digital output signal in percent:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

F

T

= 0.05 %/10 K x TD

F

Kl

= 0.1 %

F

perf

= √((0.05 %/10 K x 20 K x 1.4)

2

+ (0.1 %)

2

)

F

perf

= 0.17 %

Total deviation digital output signal in percent:

F

total

= F

perf

+ F

stab

F

stab

= 0.1 %/year x TD

F

stab

= 0.1 %/year x 1.4

F

stab

= 0.14 %

F

total

= 0.17 % + 0.14 % = 0.31 %

Total deviation digital output signal absolute:

F

total

= 0.35 % x 147 mbar/100 % = 0.46 mbar = 4.7 mm

Basic accuracy analogue output signal in percent:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

+ (F

a

)

2

)

F

T

= 0.05 %/10 K x TD

F

Kl

= 0.1 %

F

a

= 0.15 %

F

perf

= √((0.05 %/10 K x 20 K x 1.4)

2

+ (0.1 %)

2

+ (0.15 %)

2

)

F

perf

= 0.23 %

Total deviation analogue output signal in percent:

F

total

= F

perf

+ F

stab

F

stab

= 0.1 %/year x TD

F

stab

= 0.1 %/year x 1.4

F

stab

= 0.14 %

F

total

= 0.23 % + 0.14 % = 0.37 %

Example

Advertising