4 Mounting

VEGAPULS 66 • 4 … 20 mA/HART - four-wire

36520-EN-121011

Fig. 13: Agitators

Through the action of filling, stirring and other processes in the vessel,

dense foams which considerably damp the emitted signals may form

on the product surface.
If foams are causing measurement errors, the biggest possible radar

antenna should be used.
As an alternative, sensors with guided microwave can be used. These

are unaffected by foam generation and are best suited for such ap-

plications.

When using a surge pipe in a vessel, influences from vessel installa-

tions and turbulences can be excluded. Under these prerequisites,

the measurement of products with low dielectric values (ε

r

value

≥ 1.6) is possible. In very adhesive products, measurement in a surge

pipe is not recommended.

Foam generation

Measurement in a surge

pipe