VEGA VEGAPULS 66 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - four-wire User Manual

Page 64

Advertising
background image

64

11 Supplement

VEGAPULS 66 • 4 … 20 mA/HART - four-wire

36520-EN-121011

Ʋ Nominal pressure stage

PN 250

Electromechanical data - version IP 66/IP 67
Options of the cable entry

Ʋ Cable gland

M20 x 1.5 (cable: ø 5 … 9 mm)

Ʋ Cable entry

½ NPT

Ʋ Blind stopper

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Supply and signal circuit (only with low

voltage)

M12 x 1

Ʋ Indication circuit

M12 x 1

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Indicating and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Measured value indication

Ʋ Number of digits

5

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

Voltage supply
Operating voltage

Ʋ Version for low voltage

9.6 … 48 V DC, 20 … 42 V AC, 50/60 Hz

Ʋ Version for mains voltage

90 … 253 V AC, 50/60 Hz

Interpolation protection

Integrated

Advertising