11 Supplement

VEGAPULS 61 • Modbus and Levelmaster protocol

41361-EN-121011

Max. torque for NPT cable glands and Conduit tubes

Ʋ Plastic housing

10 Nm (7.376 lbf ft)

Ʋ Aluminium/Stainless steel housing

50 Nm (36.88 lbf ft)

Input variable
Measured variable

The measured quantity is the distance between process

fitting of the sensor and product surface. The reference

plane is the seal surface on the hexagon or the lower

side of the flange.

3

4 2

1

Fig. 51: Data of the input variable
1 Reference plane

2 Measured variable, max. measuring range

3 Antenna length

4 Useful measuring range

Standard electronics
Max. measuring range

35 m (114.8 ft)

Recommended measuring range

Ʋ Encapsulated antenna system

up to 10 m (32.81 ft)

Ʋ Plastic horn antenna

up to 20 m (65.62 ft)

Electronics with increased sensitivity
Max. measuring range

35 m (114.8 ft)

Recommended measuring range

Ʋ Encapsulated antenna system

up to 10 m (32.81 ft)

Ʋ Plastic horn antenna

up to 20 m (65.62 ft)

Output variable
Output

Ʋ Physical layer

Digital output signal according to standard EIA-485

Ʋ Bus specifications

Modbus Application Protocol V1.1b, Modbus over serial

line V1.02