VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol User Manual

Page 63

Advertising
background image

63

11 Supplement

VEGAPULS 61 • Modbus and Levelmaster protocol

41361-EN-121011

Deviation with bulk solids

The values depend to a great extent on the application.

Binding specifications are thus not possible.

Variables influencing measurement accuracy
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm

Additional deviation through strong, high

frequency electromagnetic fields acc. to

EN 61326

< ±50 mm

Characteristics and performance data
Measuring frequency

K-band (26 GHz technology)

Measuring cycle time

Ʋ Standard electronics approx.

450 ms

Ʋ Electronics with increased sensitivity

approx.

700 ms

Step response time

1)

≤ 3 s

Tracking speed of the measuring window

max.

1 m/min

Beam angle

2)

Ʋ Encapsulated antenna system

22°

Ʋ Plastic horn antenna

10°

Emitted HF power (depending on the parameter adjustment)

3)

Ʋ Average spectral transmission power

density

-14 dBm/MHz EIRP

Ʋ Max. spectral transmission power

density

+43 dBm/50 MHz EIRP

Ʋ Max. power density in a distance of

1 m

< 1 µW/cm²

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
For the process conditions, please also note the specifications on the type label. The lower value

always applies.
Vessel pressure

Ʋ Encapsulated antenna system

-1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psi)

Ʋ Plastic horn antenna

-1 … 2 bar (-100 … 200 kPa/-14.5 … 29.0 psig)

Ʋ Version with adapter flange from

DN 150 PP

-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)

1)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)

3)

EIRP: Equivalent Isotropic Radiated Power