VEGA VEGAPULS 65 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 54

Advertising
background image

54

11 Supplement

VEGAPULS 65 • HART and accumulator pack

40932-EN-121011

1,0 m (3.28 ft)

20 mm (0.788 in)

8 mm (0.315 in)

- 8 mm (- 0.315 in)

- 20 mm (- 0.788 in)

0

1

2

3

Fig. 38: Deviation under reference conditions
1 Reference plane

2 Antenna edge

3 Recommended measuring range

Reproducibility

≤ ±1 mm

Variables influencing measurement accuracy
Specifications apply to the HART signal and the current output
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm

Additional deviation through strong, high

frequency electromagnetic fields acc. to

EN 61326

< ±50 mm

Specifications apply also to the current output
Temperature drift - Current output

±0.03 %/10 K relating to the 16 mA span max. ±0.3 %

Deviation on the current output by ana-

logue/digital conversion

< ±15 µA

Deviation on the current output due to

strong, high frequency electromagnetic

fields acc. to EN 61326

< ±150 µA

Characteristics and performance data
Measuring frequency

C-band (6 GHz technology)

Measuring cycle time approx.

350 ms

Step response time

1)

≤ 3 s

Tracking speed of the measuring window

max.

1 m/min

Beam angle

2)

24°

Emitted HF power (depending on the parameter adjustment)

3)

Ʋ Average spectral transmission power

density

-31 dBm/MHz EIRP

1)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)

3)

EIRP: Equivalent Isotropic Radiated Power

Advertising