11 Supplement

VEGABAR 86 • 4 … 20 mA

45506-EN-131010

Ʋ Displayed value 1

Pressure in bar/mbar

Ʋ Displayed value 2

Pressure in bar/mbar

Output variable - Second current output
For details on the operating voltage see chapter "Voltage supply"
Output signal

4 … 20 mA (passive)

Range of the output signal

3.8 … 20.5 mA (default setting)

Signal resolution

0.3 µA

Failure signal current output (adjustable) Last valid measured value, ≥ 21 mA, ≤ 3.6 mA
Max. output current

21.5 mA

Starting current

≤ 10 mA for 5 ms after switching on, ≤ 3.6 mA

Load

Load resistor, see chapter "Voltage supply"

Damping (63 % of the input variable),

adjustable

0 … 999 s

Dynamic behaviour output
Run-up time approx.

10 s

Dynamic characteristics, depending on medium, temperature and, if applicable, chemical seal

90 %

100 %

10 %

t

t

T

A

t

t

S

2

1

Fig. 25: Behaviour in case of sudden change of the process variable. t

T

: dead time; t

A

: rise time; t

S

: jump response

time
1 Process variable

2 Output signal

Dead time

> 45 ms

Rise time

> 35 ms (10 … 90 %)

Step response time

> 80 ms (ti: 0 s, 10 … 90 %)

Damping (63 % of the input variable)

0 … 999 s, adjustable

Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+15 … +25 °C (+59 … +77 °F)

Ʋ Relative humidity

45 … 75 %