VEGA VEGABAR 83 4 … 20 mA - Operating Instructions User Manual

Page 27

Advertising
background image

27

6 Set up with the display and adjustment module

VEGABAR 83 • 4 … 20 mA

45033-EN-131010

Carry out the following steps in the below sequence.
You can find "Extended adjustment" in the next sub-chapter.
1. Measurement loop name

In the first menu item you assign a suitable measurement loop name.

Permitted are names with max. 19 characters.
2. Application

In this menu item you activate/deactivate the slave for the electronic

differential pressure and select the application. With process pressure

transmitters, the selection comprises process pressure and level

measurement; with suspension pressure transmitters it comprises

level measurement.
3. Units

In this menu item you determine the adjustment and temperature

units of the instrument. Depending on the selected application in the

menu item "Application", different adjustment units are available.

4. Position correction

In this menu item you compensate the influence of the installation

position of the instrument (offset) to the measured value.
5. zero adjustment

In this menu item you carry out the zero adjustment for the processs

pressure.
Enter the respective pressure value for 0 %.
6. span adjustment

In this menu item you carry out the span adjustment for the processs

pressure
Enter the respective pressure value for 100 %.

4. Position correction

In this menu item you compensate the influence of the installation

position of the instrument (offset) to the measured value.
5. Max. adjustment

In this menu item you carry out the max. adjustment for level

Presettings

Adjustment - Process

pressure measurement

Adjustment - Level meas-

urement

Advertising