3 Product description

VEGABAR 83 • 4 … 20 mA

45033-EN-131010

Measured products are gases, vapours and liquids.
VEGABAR 83 is especially suitable for applications with higher tem-

peratures and high pressures.

The process pressure acts on the sensor element via the process

diaphragm and an internal transmission liquid. The process pressure

causes a resistance change which is converted into a corresponding

output signal and outputted as measured value.
Measuring ranges up to 16 bar: piezoresistive sensor element, meas-

uring ranges up to 25 bar: strain gauge (DMS) sensor element.

1

3

4

2

Fig. 3: Configuration of the measuring system with piezoresistive sensor ele-

ment
1 Sensor element

2 Base element

3 Transmission liquid

4 Process diaphragm

With small measuring ranges or higher temperatures, the METEC

®

measuring cell is the measuring unit. It consists of the ceramic-capac-

itive CERTEC

®

measuring cell and a special, temperature-compen-

sated chemical seal system.

4

3

2

1

Fig. 4: Configuration of the METEC

®

measuring cell

1 Process diaphragm

2 Isolating liquid

3 FeNi adapter

4 CERTEC

®

measuring cell

The measuring cell design depends on the selected pressure type.