VEGA VEGAFLEX 81 4 … 20 mA_HART - four-wire User Manual

Page 73

Advertising
background image

73

11 Supplement

VEGAFLEX 81 • 4 … 20 mA/HART - four-wire

41825-EN-130314

Ʋ Cable probes

ε

r

≥ 1.6

Ʋ Rod probes

ε

r

≥ 1.6

2

4

5

3

2

4

5

3

1

1

Fig. 36: Measuring ranges - VEGAFLEX 81
1 Reference plane

2 Probe length L

3 Measuring range (default setting refers to the measuring range in water)

4 Upper dead band (see diagrams under Accuracy - grey section)

5 Lower dead band (see diagrams under Accuracy - grey section)

Output variable
Output signals

4 … 20 mA/HART - active; 4 … 20 mA/HART - passive

Terminal voltage passive

9 … 30 V DC

Shortcircuit protection

Available

Potential separation

Available

Signal resolution

0.3 µA

Failure signal current output (adjustable) Last valid measured value, ≥ 21.0 mA, ≤ 3.6 mA
Max. output current

22 mA

Starting current

≤ 3.6 mA

Load (4 … 20 mA/HART - active)

< 500 Ω

Damping (63 % of the input variable)

0 … 999 s, adjustable

HART output values according to HART 7 (default setting)

1)

Ʋ First HART value (PV)

Linearised percentage value, level

Ʋ Second HART value (SV)

Distance to the level

1)

The output values can be assigned individually