10 Diagnostics and service

VEGAFLEX 81 • 4 … 20 mA/HART two-wire

44216-EN-130823

Code
Text mes-

sage

Cause

Rectification

C700
Simulation ac-

tive

– A simulation is active

– Finish simulation

– Wait for the automatic end

after 60 mins.

C701
Parameter

verification

– Parameter verification was

interrupted

– Finish parameter verifica-

tion

The following table shows the error codes and text messages in the

status message "Out of specification" and provides information on

causes as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

S600
Unpermissi-

ble electronics

temperature

– Temperature of the pro-

cessing electronics in the

non-specified section

– Check ambient temperature

– Isolate electronics

– Use instrument with higher

temperature range

S601
Overfilling

– Level echo in the close

range not available

– Reduce level

– 100 % adjustment: Increase

value

– Check mounting socket

– Remove possible interfering

signals in the close range

– Use coaxial probe

S602
Level with-

in the search

range, com-

pensation

echo

– Compensation echo super-

imposed by medium

– 100 % adjustment: Increase

value

S603
Impermissi-

ble operating

voltage

– Operating voltage below

specified range

– Check electrical connection

– if necessary, increase

operating voltage

The following table shows the error codes and text messages in the

status message "Maintenance" and provides information on causes

as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

M500
Error in the

delivery sta-

tus

– With the reset to delivery

status, the data could not

be restored

– Repeat reset

– Load XML file with sensor

data into the sensor

Out of specification

Maintenance