VEGA VEGAFLEX 81 4 … 20 mA_HART two-wire Coax probe With SIL qualification User Manual

Page 80

Advertising
background image

80

12 Supplement

VEGAFLEX 81 • 4 … 20 mA/HART two-wire

44216-EN-130823

1

2

0°C

(32°F)

-40°C

(-104°F)

65°C

(149°F)

80°C

(176°F)

150°C

(302°F)

130°C

(266°F)

3

4

-60°C

(-140°F)

Fig. 43: Ambient temperature - process temperature, elastomer-free version for ammonia applications
1 Ambient temperature

2 Process temperature (depending on the seal material)

3 Maximum permissible temperature - standard

4 Limited temperature range - plastic housing and stainless steel housing, electropolished

Viscosity - dynamic

0.1 … 500 mPa s (requirement: with density 1)

Vibration resistance

Ʋ Instrument housing

4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

Ʋ Coax probe

1 g with 5 … 200 Hz according EN 60068-2-6 (vibration

at resonance) with tube length 50 cm (19.69 in)

Shock resistance

Ʋ Instrument housing

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Ʋ Coax probe

25 g, 6 ms according to EN 60068-2-27 (mechanical

shock) with tube length 50 cm (19.69 in)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Ʋ Cable entry

½ NPT

Ʋ Blind plug

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

Plug M12 x 1, according to ISO 4400, Harting HAN,

7/8" FF

Ʋ Indication circuit

M12 x 1 plug

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Advertising