6 Set up with the display and adjustment module

SOLITRAC 31 • Profibus PA

43834-EN-131119

The following reset functions are available:
Basic settings: Resetting of the parameter adjustments to default

values at the time of shipment. Order-specific settings are deleted.
Default settings: Resetting of the parameter adjustment like under

"Basic settings". In addition, special parameters are reset to default

values. Order-specific settings are deleted.
Peak values of measured value: Resetting of the parameter adjust-

ments in the menu item "Setup" to the default values of the respective

instrument. Order-specific settings remain but are not taken over into

the current parameters.
Peak values of temperature: Resetting of the measured min. and

max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The

values apply for the application "Level". First of all you have to select

the application.
Depending on the instrument version, not all menu items may be

available or they may be differently assigned:

Menu

Menu item

Default value

Setup

Sensor address

126

Measurement loop name

Sensor

Isotope

Cs-137

Application

Level

Background radiation

0 ct/s

Unit of the process value

%

Temperature unit

° C

Adjustment min.

Min. process value = 0 %

Adjustment max.

Max. process value = 100 %

Linearization

0 ct/s = 100 %
90000 ct/s = 0 %

AI FB1 Channel

Process value

AI FB1 scaling unit

%

AI FB1 scaling

0 % = 0 %
100 % = 100 %

AI FB1 damping

0 s

Damping

60 s

Real value correction

0

Reference value - Relay

None

Block adjustment