7 software update – VEGA VEGAFLEX 83 Foundation Fieldbus Polished version User Manual
Page 68
68
9 Diagnostics and service
VEGAFLEX 83 • Foundation Fieldbus
44224-EN-130910
Fig. 51: Exchanging the measuring rod
Information:
Please keep the stated torque so that the max. tensile strength
remains.
7. Enter new probe length and if necessary the new probe type and
then carry out a fresh adjustment (see "Setup procedure, Carry-
ing out min. adjustment - Carrying out max. adjustment").
9.7 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedur
"Software".
You can find information about the installation in the download file.