4 mounting, 1 general instructions, 2 mounting preparations – VEGA VEGAFLEX series 80 Electronics module User Manual

Page 7: 3 mounting steps

Advertising
background image

7

4 Mounting

Electronics module • VEGAFLEX series 80

43656-EN-121113

4 Mounting

4.1 General instructions

The new electronics module must be loaded with the order and fac-

tory data of the sensor. These are the options:

At the factory by VEGA

Or on site by the user

Information:

When loading on site, first of all the order data must be downloaded

from the Internet (see "Setup").
In both cases, the sensor serial number is necessary. The serial num-

bers are stated on the type label of the instrument, inside the housing

or on the delivery note.
Caution:

The order data and company data contain important presettings for

the sensor. A reliable operation and a correction of the measurement

is not possible without these data.

Warning:

Switch off voltage supply before starting the mounting procedure. The

replacement electronics must only be mounted in idle condition.

Non-observance will damage the electronics!

The safety-relevant data for explosion protection of the electronics

module supplied later can deviate from the data of the previously

installed electronics module. If the electronics modue supplied later,

is installed in a certified, explosion-protected sensor, it is absolutely

necessary to observe the instrument "Safety instructions" attached to

the supplied electronics module! The attached label must be applied

on the sensor housing.

4.2 Mounting preparations

The electronics modules are adapted to the respective sensor. First

of all check by means of the following lists if you are using a suitable

electronics module.

FX-E.80H for version 4 … 20 mA/HART (two-wire/four-wire/Mod-

bus)

FX-E.80A for version 4 … 20 mA/HART with SIL qualification

FX-E.80P for version Profibus PA

FX-E.80F for version Foundation Fieldbus

4.3 Mounting steps

The electronics module is installed in the electronics compartment.

The following illustrations show the respective position of the elec-

tronics compartment in the single or double chamber housing.

Sensor serial number

Assignment

Mounting steps

Advertising