VEGA VEGABAR 86 Save sensor - Operating Instructions User Manual

Page 46

Advertising
background image

46

9 Supplement

VEGABAR 86 • Slave sensor for electronic differential pressure

45052-EN-131011

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Ʋ Cable entry

½ NPT

Ʋ Blind plug

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

Plug M12 x 1, according to ISO 4400, Harting HAN,

7/8" FF

Ʋ Indication circuit

M12 x 1 plug

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Electromechanical data - version IP 68 (25 bar)
Connection cable between IP 68 instrument and external housing

Ʋ Configuration

four wires, one suspension wire, one breather capillary,

screen braiding, metal foil, mantle

Ʋ Wire cross-section

0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

Ʋ Standard length

5 m (16.40 ft)

Ʋ Max. length

180 m (590.5 ft)

Ʋ Min. bending radius at 25 °C/77 °F

25 mm (0.985 in)

Ʋ Diameter approx.

8 mm (0.315 in)

Ʋ Colour

Blue

Cable entry/plug

4)

Ʋ External housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Spring-loaded terminals for wire cross-

section up to

2.5 mm² (AWG 14)

Voltage supply, interface to the master sensor
Voltage supply

Through the master sensor

Data transmission

digital (I²C-Bus)

Configuration, connection cable

4-wire, screened

Cable length max.

25 m

Potential connections in the instrument
Electronics

Not non-floating

4)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.

Advertising