4 Mounting

External housing • VEGAFLEX 80 series

46802-EN-131120

4 Mounting

4.1 General instructions

In the following cases, we recommend using an instrument version

with external housing:

if the standard housing is too big

if strong vibrations can damage the electronics

In Ex applications, only a housing with appropriate Ex approval must

be used.

4.2 Mounting preparations

The following tools are required for mounting the external housing.

Instrument housing:

Hexagon spanner size 4

Fork wrench, wrench size 19

Sensor housing:

Fork wrench, wrench size 8

Fork wrench, wrench size 19

We recommend using further material when mounting the wall mount-

ing plate.

4 screws, depending on the mounting surface

4.3 Mounting steps - instrument housing

1. Mark the holes according to the following drilling template
2. Depending on the mounting surface, fasten the wall mounting

plate with 4 screws

90 mm

(3.54")

R 3,5 mm

(0.14"

)

3mm

(0.12")

70 mm

(2.76")

8 mm

(0.32")

93 mm (3.66"

)

110 mm (4.33"

)

Fig. 2: Hole pattern - wall mounting plate (external housing)