9

5 Connect the sensor to the external housing

External housing • VEGAFLEX 80 series

46802-EN-131120

5 Connect the sensor to the external

housing

5.1 Preparing the connection

Follow the instructions in the operating instructions manual of the

sensor.

5.2 Connection procedure

Note:

Only use the corresponding HF connection cable in the correct

length.
Make sure that the markings (serial no.) correspond on the individual

parts (sensor housing, instrument housing and HF connection cable).
The confectioned HF connection cable has a straigth N-plug on one

end and on the other end an N-angled plug. Both plugs can be con-

nected to any side.
1. Remove the protective cover on the plug connector of the sensor

housing.

2. Insert the plug (straight plug or angle plug) of the HF connection

cable on the sensor housing and hand-screw the hexagon nut.

3. Tighten the hexagon nut of the plug with the fork wrench (SW 19)

with max. 3 Nm (2.212 lbf ft).

4. Secure the hexagon nut of the plug with the securing strap. For

this purpose, place the securing strap on one surface of the hexa-

gon nut and tighten it with an Allen key (size 4).

3

1

2

Fig. 3: Place the securing strap to the sensor housing and tighten it
1 Safety catch

2 Locking screw

3 Hexagon nut of the plug

5. Remove the protective cover on the plug connector of the instru-

ment housing.

6. Insert the plug (straight plug or angle plug) of the HF connection

cable on the instrument housing and hand-screw the hexagon

nut.

7. Tighten the hexagon nut of the plug with the fork wrench (SW 19)

with max. 3 Nm (2.212 lbf ft).