Charging the battery requires a common 9VDC adapter (positive pole located inside the female connector) supplied with the analyzer
and a convenience outlet. The analyzer’s charging circuit accepts 9VDC from any standard AC 110V or 220V adapter (specified when
the order is placed). The electronic design enables the analyzer to remain fully operable during the 8-10 hour charging cycle.

When power is applied, the software performs several diagnostic system status checks termed “START-UP TEST” as illustrated below.
Note: The analyzer display defaults to the sampling mode when 30 seconds elapses without user interface.

START-UP TEST


ELECTRONICS – PASS
LOOP POWER – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS

93.3

%

AUTO SAMPLING

100% RANGE

24.5 C 100 KPA

Menu Navigation


The four (4) pushbuttons located on the front of the portable oxygen purity analyzer operate the micro-processor:
1. Green ENTER (select)
2. Yellow UP ARROW
3. Yellow DOWN ARROW
4. Blue MENU (escape)

Main Menu


Access the MAIN MENU by pressing the MENU key:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATE

24.5 C 100 KPA


Range Selection:


The GPR-3500MO portable oxygen purity analyzer is equipped with a single 0-100% measuring range (see specification); accordingly,
there is no difference between auto and manual sampling.




Start-Up is complete . . .