Advanced Instruments Inc.



32

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS



>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO


5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.

6. Press the ENTER key to select the highlighted menu option.

7. The following display appears:

100.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN


8. Enter the adjustment value as follows . . .
9. Selecting digits of the adjustment value to change - Press the ENTER key to advance the underline cursor right or press the MENU

key to advance the underline cursor left to reach to the desired digit.

10. Changing the digits of the adjustment value . . . Press the UP/DOWN ARROW keys to change a digit of the adjustment value.
11. Repeat steps 9 and 10 until the complete span value has been entered and the following display appears.

064.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN


12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
14. Note: After the initial adjustment is made, check the actual signal output at the PLC and if necessary calculate a further fine

tuning adjustment value as illustrated above.

15. Enter subsequent fine tuning adjustments as described above.

075.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN


16. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
17. The system returns to the SAMPLING mode.