8 Troubleshooting


Symptom

Possible Cause

Recommended Action

Reading does not reflect
expected values

Sensor was not calibrated at the pressure,
flow rate and temperature anticipated in the
sample gas stream

Recalibrate the analyzer

Oxygen reading drifts toward
zero

Indication sensor is nearing the end of its
useful life

Replace sensor, see section 6 Maintenance


Slow response time

Erratic oxygen reading


No oxygen reading


Liquid covering sensing membrane

Presence of interference gases
Unauthorized maintenance

Defective electrical connection
Sensor failure


Gently remove with damp lint free towel.

Consult factory


Consult factory
Replace sensor and/or consult factory


High oxygen reading


Inadequate control of flow rate
Abnormality in span gas


See Pressure and Flow
Qualify source