Confi guration – Xylem HandyLab 780 User Manual

Page 54

Advertising
background image

54

pH

Confi guration

pH Confi guration

1) In measuring mode, press the Menu softkey.
2) Select “Confi guration” and confi rm by pressing enter.
3) Make the desired adjustments.
The following table gives you an overview.
Factory settings are shown in bold print.

*) “+” indicates that submenus can be opened by pressing enter.
**) Parameter can be confi gured using HandyLab Pilot software.

“Confi guration” menu selection – part 1

Language

Deutsch

| English | Español | Italiano | Français |

Portu-

guês

Auto-off

Off

| 5 min | 10 min | 30 min | 60 min

Temperature

°C

| °F

+ pH sensor

*)

Wear

On

| Off

+ Calibration

*)

Calibration timer

Off

| On

Interval

On: 00 … 99 days

Calibration mode

AutoCal

| Manual | Data entry

Calibration points

Auto

| 1-point | 2-point | 3-point

Buffer set

Mettler-Toledo
Knick CaliMat
Ciba
NIST technical
NIST standard
Hach
WTW
Hamilton
Reagecon
DIN 19267
User buffer 1

**)

.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75

MemoLog

Off

| On

TAG

Off

| On

-01-

(De-

fault)

NIST standard

-02-

NIST technical

-03-

SI Analytics

technical

-04-

Knick Calimat

-05-

Mettler Toledo

-06-

HACH

-07-

Ciba (94)

-08-

Hamilton

-09-

Reagecon

-10-

DIN 19267

1,679/4,006/6,865/9,180

1,68/4,00/7,00/10,01/12,46

2,00/4,00/7,00/10,00
2,00/4,00/7,00/9,00/12,00
2,00/4,01/7,00/9,21
4,01/7,00/10,01/12,00
2,06/4,00/7,00/10,00
2,00/4,01/7,00/10,01/12,00
2,00/4,00/7,00/9,00/12,00
1,09/4,65/6,79/9,23/12,75

User-Puffer 1

**)

Advertising