YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual

Page 23

Advertising
background image

MIQ/CHV PLUS

Settings

4 - 7

ba76036e01

01/2012

Settings

* With short cleaning intervals, the adjustable values for the Cleaning
duration
and Adjustment time are limited. The following values apply:

Note
With this, the cleaning times are fixed. They only change when the
reference time is changed.

Setting

Selection/Values

Explanation

Reference time (h)

0 ... 23 h

Time at which a
cleaning cycle is
started. Further
cleaning cycles will
be performed at the
times specified by the
cleaning interval.

Reference time
(min)

0 ... 60 min

Interval unit

1 .. 7 d
1 .. 24 h
5 .. 60 min

Selection of range
and unit for the
Cleaning interval.

Cleaning interval

1/2/3/4/5/6/7 d

or:
1/2/3/4/6/8/12/24 h

or:
5/10/15/20/30/60 min

Repeat interval for
the cleaning function:
Time between the
start time of a
cleaning cycle and
the start time of the
next cleaning cycle*.

Cleaning duration

0 ... 300 s

Duration of the
cleaning

Adjustment time

0 ... 900 s

Time extension to
allow the sensor to
adjust to the test
sample after
cleaning.

Cleaning interval

Cleaning duration

Adjustment time

10 min

max. 60 s

max. 120 s

20 min

max. 180 s

max. 300 s