Technical data

MIQ/CHV PLUS

7 - 4

ba76036e01

01/2012

Valve circuits

Warning

The valve circuit must not supply any unauthorized voltages or
currents. It has to be made sure that the circuit at any time meets
all requirements of a Limited circuit
or Limited Power as well as of
SELV
(Safety Extra Low Voltage). These include the following
limiting value specifications:

AC voltage: max. 30 V effective / 42.4 V peak

DC voltage: max. 60 V

Current limit: max. 8 A

Power output limitation: max. 150 VA

Compressed air

Switching voltage

Approx. 12 V

Max. switching current

Approx. 70 mA

Required air quality

Dry, dust-free and oil-free

Operating pressure

Max. 7x10

5

Pa (7 bar) absolute

Connections

6 mm hose nozzles