Xylem EX100/200-Series ELECTROMAGNETIC FLOW SENSORS User Manual

Page 14

Advertising
background image

Min 0.28

Max 20.0

FLOW RATES (IN GALLONS PER MINUTE)

Feet Per

Second

NOMINAL PIPE SIZE

3” 4” 6” 8” 10” 12” 14” 16” 18” 20” 24” 30” 36” 48”

6

11

25

44

69

99

134

175

222

274

395

617

888

1580

440

783

1,762

3,133

4,895

7,050 9,596 12,533 15,863 19,584 28,200 44,064 63,452 112,804

OPERATION & MAINTENANCE

Page 12

Filtering.

The software of the EX100/200-Series filters out

electrical noise and averages sudden variations in the flow
to smooth the output. It takes a matter of seconds for the
flow sensor to get up to full output when it is powered up or
when flow begins.

Electrode Coating.

Grease or other adhering, non-conductive

materials can stop flow detection if the electrodes become
heavily coated. To clean the electrodes, remove the sensor
from the pipe and gently scrub the electrodes (three silver
bumps) on the reading face of the flow sensor. A mild soap
(dishwashing liquid for example) can be used to aid the
cleaning process.

Calibration (“K-Factor”).

In order to properly process pulses

from the flow sensor, a number must be entered into the
control to which the sensor is connected. This number,
called the K-factor, is the number of pulses the sensor puts
out per unit of fluid passing through the pipe. It is normally
provided for Seametrics sensors in pulses per gallon, and
can be ascertained by using the “K-Factor Calculator” on the
Seametrics website. These numbers are based on extensive
testing, which has shown close agreement among different
EX sensors in the same installation. Typically, most K-factor
error can be attributed to installation variables, such as depth
setting and fitting configuration.

To perform the adjustment, after determining that there is a
full pipe with no flow, short between the two pins marked “Zero
Adjust”. A red LED light will come on for approximately 50 sec-
onds and then go out. The zero adjustment is completed.

Minimum Flow.

As with any other flow sensor, there is a rate

below which the EX100-Series sensor cannot read. Check
the table below for the minimum flow rate detectable by the
sensor for a given pipe size.

Presence of Flow Indication.

To assist in troubleshooting,

the “Status LED” has two blinking modes in normal operation.
When there is no flow detectable by the meter (below
minimum threshold) the LED blinks every 8.0 seconds. When
there is detectable flow, the same indicator blinks every 3.0
seconds (Pulses are being output when indicator is blinking
every 3 seconds).

Zero Adjustment.

When the EX100/200-Series meter is pow-

ered up and there is no flow, there should be no output pulses
(or, if connected to the FT420, flow rate should read “0”). If
there are pulses, it may be necessary to adjust the flow meter
under no-flow conditions after it has been installed. This should

only be done if the indicated flow is low, near the lower cutoff.

30Vdc

Max. 6mA

- +

- +

12-24Vdc

3

4

5

6

2

1

Power

Forward

Output

Status

LED

Zero

Adjust

Pins

Zero

Adjust

Pins

Status

LED

Using the K-Factor Calculator:

1) Go to the Seametrics webpage, EX K-factor calculator

(http://www.seametrics.com/node/223)

*

2) Select your meter

3) Choose units

4) Input external pipe diameter (measure) and wall

thickness (measure, or look-up in Table 2 on page 5

of this manual)

5) Press Calculate to determine your K-factor

(NOTE: Dimension D is also calculated.)

6) Enter your K-factor into your controller

*For pipe sizes larger than 50”, please consult factory.

Advertising